Gastec 600 Injection Type Gas Measurement System Propane

In No.600 System, samples are injected into detector tubes by using a syringe in order to measure gases in very high concentrations or slow in chemical reactions.

Description

In No.600 System, samples are injected into detector tubes by using a syringe in order to measure gases in very high concentrations or slow in chemical reactions.

Brand

Gastec

GASTEC

Although, technology is always evolving, one demand on our products never changes: the demand for customer safety. Whenever we contribute to customer safety, we are presented with an opportunity to develop further.

GASTEC holds this corporate creed of safety and development in high esteem, as it strives to grow while always putting customer safety first.

Features

  • Compact, lightweight, and versatile(can be used anywhere)
  • Easy direct measuring and reading
  • No sampling pump necessary

Applications

  1. Break off both ends of a detector tube (No.100B for No.601 syringe or No.2HT for No.611 syringe) by using the tube tip cutter.
  2. Attach the syringe inlet to the syringe (No.601 or No.611) and collect 20ml of sample into the syringe by pulling its piston.
  3. Insert the tube securely into the syringe inlet with the arrow pointing away from the syringe. Inject the sample collected in the syringe into the detector tube in 2 minutes (for No.100B) or 20 seconds (for No.2HT), and read the indication.
  4. Please see the product manual for details.

Specifications

Substance to be measuredTube no. & nameMeasuring range
(ppm)
Scale range
(ppm)
No. of
tests
per box
Detector tube typeNotes
Propane100BPropane0.1-2.0%0.1-2.0%10Injection typeNo.600 (Propane)

Couldn’t find what you were looking for?

Contact us with your specification for assessment and availability of other products configurations.